Bơm lấy mẫu khí cá nhân_50 ml/ phút ÷ 12 lít/phút
Model: Escort ELF
Code: 805558
Hãng sản xuất: Zefon Escort – Anh
- Escort ELF air sampling pumps feature a state of the art electronic laminar flow sensor to provide a constant flow control with ± 2.5% regulation of flow rate (from 1-3 lpm) with automatic compensation for changes in battery voltage, temperature, altitude, and sample load. An internal secondary standard calibrates the pump continuously and needs to checked against a primary standard only once per month (or every 200 hours for coal mine dust sampling).
Features:
- Patented electronic laminar flow control.
- Internal secondary standard calibrates the pump continuously, maintaining OSHA compliance.
- Check against a primary standard only once a month or every 200 hours for coal mine dust sampling.
- Inlet filter traps moisture and particular matter.
- Stainless steel-filled plastic case provides EMI/RFI protection.
- Rated IP66 for resistance to water and dust ingress.
- Exceptionally quiet operation.
- Flow-Fault Indicator shows any sample inlet blockage.
- Run/Hold feature allows “time out” during sampling process.
- Elapsed time readout and low-battery warning.
- Low flow and dual tube sampling available with Gemini Twin-Port Sampler.
- 2 Year Warranty.
Approvals:
- NIOSH certified for coal mine dust sampling (TC-74-030)
- MSHA certified as intrinsically safe for underground use (Approval # 2G-3924-1).
- UL approved as intrinsically safe for use in hazardous locations, Class I, Groups A, B, C, D; Class II, Groups E, F and G; Class III, Division I locations.
- ATEX Approved models available. Contact Zefon or your Zefon distributor for details.
Specifications:
Operating Flow Control
|
Volumetric flow rate held within ± 2.5% of set-point over the 1-to-3 lpm operating range (± 5% to 0.5 lpm), with automatic compensation for battery voltage, altitude, temperature and sample load changes.
|
Flow Rate
|
Flow rate adjustable between 0.5 and 3 lpm; lower flow rates attainable (to 1 mlpm) with optional Gemini Twin-Port Sampling accessory.
|
Flow Indication
|
Liquid Crystal Display provides 0.01 lpm resolution.
|
Operating Range
|
30 inches of water load up to 2 lpm and 10 inches up to 3 lpm.
|
Flow Blockage Detection
|
Flow-Fault LED comes on immediately when block is detected. Pump shuts down within 90 seconds if block is not cleared.
|
Elapsed Time Readout
|
To 999 minutes in increments of 1 minute. Holds the last reading after either fl ow blockage or low battery shutdown and when pump is off or in hold mode.
|
Operating Time
|
Varies with flow rate and sampling device loading.
|
Operating Temperature Limits
|
32°-113°F (0°-45°C)
|
Physical Weight
|
23 oz. (652 gm), with battery pack.
|
Dimensions
|
2.25” D x 4.25” H x 4” W
(5.7 cm x 10.8 cm x 10.2 cm), with battery pack. |
Power Supply
|
4.8-volt battery pack of four NiMH cells. (Units Mfg before Oct 2011 contain NiCd cells)
|
Battery Pack Capacity
|
2.8 amp hours (NiMH version), 2.4 amp hours (NiCD version)
|
Battery Pack Recharge Time
|
14-16 hours (overnight) with Omega™ Charger.
|
Typical Battery Pack Life
|
300 or more charging cycles
|